Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Randox Laboratories Ltd. recalls Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other method may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.
Products
  • Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426
Where it was sold

US Nationwide distribution in the states of GA, MA, WV including PR.

States: Georgia, Massachusetts, Puerto Rico, West Virginia

Barcodes
05055273204544
Units affected
6 units
About this recall
Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426