Raye's Inc. recalls SW Bari Lift & Transfer
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.
- Products
-
- SW Bari Lift & Transfer
- Company
- Raye's Inc.
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Australia, UK.
Nationwide
- Barcodes
00845699001040- Units affected
- 1,149 units
- About this recall
- SW Bari Lift & Transfer, Model Number 38060000