Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Siemens Medical Solutions USA, Inc. recalls ACUSON Juniper Diagnostic Ultrasound System

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.
Products
  • ACUSON Juniper Diagnostic Ultrasound System
Where it was sold

Worldwide distribution - US Nationwide including in the states of AL, AR, CA, CO , CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA , MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SD, TN, TX, UT, VA, WA, WI, WY and the countries of BG, BW, BA, TR, AR, CL, CN, FR, SK, CD, BH, PH, AM, CR, AF, BE, AE, AZ, AO, KR, RO, HU, HR, DK, AU, DZ, CY, PL, BO, AL, NL, CI, BR, CZ, DE, AT, ES, CO, BD, CH, IT, CA, CM, DO, EC, EG, ER, ET, FI, FO, GB, GE, GR, GT, GY, HK, HN, ID, IE, IL, IN, IQ, IR, JO, JP, KE, KW, KZ, LB, LT, LU, LV, LY, MA, MD, ME, MK, MM, MN, MU, MW, MX, MY, NG, NO, NP, NZ, OM, PA, PE, PG, PK, PT, PY, QA, RS, RU, SA, SE, SG, SI, SM, SN, SV, SZ, TH, TN, TT, TW, TZ, UA, UG, UY, UZ, VN, XK, YE, ZA, ZM, ZW.

Nationwide

Units affected
5090
About this recall
ACUSON Juniper Diagnostic Ultrasound System, REF: 11335791, and ACUSON Juniper Select Diagnostic Ultrasound System, REF: 11653093, running software version VA10x with DICOM SR Cardiac option, from VB10A to VB10G, VB11A, VB11B, VB30B and VB30C