Siemens Medical Solutions USA, Inc. recalls ACUSON Juniper Diagnostic Ultrasound System
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.
- Products
-
- ACUSON Juniper Diagnostic Ultrasound System
- Where it was sold
-
Worldwide distribution - US Nationwide including in the states of AL, AR, CA, CO , CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA , MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SD, TN, TX, UT, VA, WA, WI, WY and the countries of BG, BW, BA, TR, AR, CL, CN, FR, SK, CD, BH, PH, AM, CR, AF, BE, AE, AZ, AO, KR, RO, HU, HR, DK, AU, DZ, CY, PL, BO, AL, NL, CI, BR, CZ, DE, AT, ES, CO, BD, CH, IT, CA, CM, DO, EC, EG, ER, ET, FI, FO, GB, GE, GR, GT, GY, HK, HN, ID, IE, IL, IN, IQ, IR, JO, JP, KE, KW, KZ, LB, LT, LU, LV, LY, MA, MD, ME, MK, MM, MN, MU, MW, MX, MY, NG, NO, NP, NZ, OM, PA, PE, PG, PK, PT, PY, QA, RS, RU, SA, SE, SG, SI, SM, SN, SV, SZ, TH, TN, TT, TW, TZ, UA, UG, UY, UZ, VN, XK, YE, ZA, ZM, ZW.
Nationwide
- Units affected
- 5090
- About this recall
- ACUSON Juniper Diagnostic Ultrasound System, REF: 11335791, and ACUSON Juniper Select Diagnostic Ultrasound System, REF: 11653093, running software version VA10x with DICOM SR Cardiac option, from VB10A to VB10G, VB11A, VB11B, VB30B and VB30C