Siemens Medical Solutions USA, Inc recalls
Siemens Medical Solutions USA, Inc is named on 24 recalls in Recall Relay's records since November 2023, from FDA (69 recalled products: FDA lists each product as its own recall); the most recent was June 8, 2026.
All 69 recalled products, newest first.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls ARTIS icono ceiling. Model Number: 11328100.Potential issue with the Display ceiling suspension (DCS Fix XL) that holds the monitor in place. Over time some screws that keep the monitor attached to the suspension may gradually loosen.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls ARTIS icono ceiling. Model Number: 11328100.Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls ARTIS icono biplane. Model Number: 11327600.Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls ARTIS icono floor. Model Number: 11327700.Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls ARTIS Pheno VE30A and VE40ADuring 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls ARTIS pheno and ARTIS Icono biplaneDuring patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system settings and imaging conditions, as well as the applicable regulatory requirements.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls ARTIS phenoDuring patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls ARTIS Icono biplaneDuring patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix T; Siemens Material Number (SMN): 11574003; With SofOrtho images acquired with preset automatic horizontal flip are not flipped and when acquiring single images, there is the possibility that an image of a previous patient may be processed.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls Interventional Fluoroscopic X-Ray System: LUMINOS Q.namix R; Siemens Material Number (SMN): 11574002; With SofOrtho images acquired with preset automatic horizontal flip are not flipped and when acquiring single images, there is the possibility that an image of a previous patient may be processed.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls NAEOTOM Alpha.Prime Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls NAEOTOM Alpha Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPTo remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls SOMATOM X.ceed Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASTo remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls NAEOTOM Alpha.Pro Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CTTo remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls LUMINOS Lotus MaxA potential issue associated with LUMINOS Lotus Max and Luminos dRF Max systems with software version VF11 with bucky wall stand for mobile detectors and without ceiling stand .When switching quickly between organ programs (OGPs) the collimation may be incorrect.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls MAMMOMAT Revelation;There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls MAMMOMAT Fusion;There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls MAMMOMAT Inspiration;There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
-
Class Iofficial
Siemens Medical Solutions USA, Inc recalls MAGNETOM Cima.X Upgrade. Model Number: 11689304.There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
Class Iofficial
Siemens Medical Solutions USA, Inc recalls MAGNETOM Verio Dot. Model Number: 10684333.There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
Class Iofficial
Siemens Medical Solutions USA, Inc recalls BIOGRAPH One (DE). Model Number: 11689172.There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
Class Iofficial
Siemens Medical Solutions USA, Inc recalls MAGNETOM Prisma. Model Number: 10849582.There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
Class Iofficial
Siemens Medical Solutions USA, Inc recalls MAGNETOM Connectom.X. Model Number: 11371480.There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
Class Iofficial
Siemens Medical Solutions USA, Inc recalls MAGNETOM Verio Dot Upgrade. Model Number: 10684334.There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
Class Iofficial
Siemens Medical Solutions USA, Inc recalls MAGNETOM Cima.X (DE). Model Number: 11647158.There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
Class Iofficial
Siemens Medical Solutions USA, Inc recalls MAGNETOM Skyra fit. Model Number: 10849580.There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
Class Iofficial
Siemens Medical Solutions USA, Inc recalls MAGNETOM Vida Fit. Model Number: 11410481.There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
Class Iofficial
Siemens Medical Solutions USA, Inc recalls MAGNETOM Vida (DE). Model Number: 11060815. 510(k): K170396There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
Class Iofficial
Siemens Medical Solutions USA, Inc recalls MAGNETOM Skyra Fit BioMatrix. Model Number: 11516217.There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
Class Iofficial
Siemens Medical Solutions USA, Inc recalls MAGNETOM Spectra as the following: 1. MAGNETOM Spectra. Model Number: 10837643. 2. MAGNETOM Spectra (CN). ModeThere is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
Class Iofficial
Siemens Medical Solutions USA, Inc recalls MAGNETOM Lumina (DE). Model Number: 11344916.There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
Class Iofficial
Siemens Medical Solutions USA, Inc recalls MAGNETOM Verio. Model Number: 10276755.There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
Class Iofficial
Siemens Medical Solutions USA, Inc recalls Biograph mMR. Model Number: 10433372.There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
Class Iofficial
Siemens Medical Solutions USA, Inc recalls MAGNETOM Skyra (DE). Model Number: 10432915. 510(k) Numbers: K101347There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls Artis Pheno. Image-Intensified Flouroscopic X-Ray System.Limited system movements after startup .
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls UROSKOP Omnia. Model Number: 10094910A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls UROSKOP Omnia Max. Model Number: 10762473A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and intIn very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls interventional fluoroscopic x-ray systemIn very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls interventional fluoroscopic x-ray systemIn very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls interventional fluoroscopic x-ray systemIn very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls ARTIS One Angiographic X-Ray SystemA potential issue with ARTIS One systems was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose-rate may be higher than intended by the user.
-
Class IIofficial
Siemens Medical Solutions USA, Inc. recalls ACUSON Redwood 1.0 and 2.0 Diagnostic Ultrasound SystemsIf ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management
-
Class IIofficial
Siemens Medical Solutions USA, Inc. recalls ACUSON Maple 1.0 Diagnostic Ultrasound SystemIf ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management
-
Class IIofficial
Siemens Medical Solutions USA, Inc. recalls ACUSON Juniper 1.0If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls Sensis Vibe systems with software version VD15B-Intended to be used as a diagnostic and administrative tool suIf the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possibility exists that the application could crash.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls Sensis Vibe systems with software version VD15B in combination with AXIOM Sensis Post Processing Workstation-IIf the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possibility exists that the application could crash.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls Sensis Vibe systems with software version VD15B in combination with VM VIRTUAL SERVER Model: 10765502-IntendedIf the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possibility exists that the application could crash.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls Sensis Vibe systems with software version VD15B in combination with Sensis High End Server -Intended to be useIf the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possibility exists that the application could crash.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls Multitom Rax -Intended to visualize anatomical structures by converting an x-ray pattern into a visible imageThe support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls Luminos dRF Max- Intended to visualize anatomical structures by converting an x-ray pattern into a visible imaThe support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls Luminos Agile Max -Intended to visualize anatomical structures by converting an x-ray pattern into a visible iThe support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls LUMINOS Lotus Max -Intended to visualize anatomical structures by converting an x-ray pattern into a visible iThe support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls ZE for Vertix Solitaire - The AXIOM Vertix Solitaire is a component and has been specially designed for examinShort circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls AXIOM MULTIX MT - Intended Radiographic X ray Table is a patient table used in conjunction with the Siemens MuShort circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls AXIOM MULTIX MT - Intended Radiographic X ray Table is a patient table used in conjunction with the Siemens MuShort circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls AXIOM Vertix Solitaire - The Vertix MD Trauma has been specially designed for examining emergency and accidentShort circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls AXIOM MULTIX MP - Intended Radiographic X ray Table is a patient table used in conjunction with the Siemens MuShort circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls MULTIX TOP - Intended Use: The Multix TOP I PRO Radiographic X ray Table is a patient table used in conjunctioShort circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls AXIOM MULTIX MP - Inended Radiographic X ray Table is a patient table used in conjunction with the Siemens MulShort circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls MULTIX TOP - Intended Use: The Multix TOP I PRO Radiographic X ray Table is a patient table used in conjunctioShort circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls AXIOM Vertix MD - The Vertix MD Trauma has been specially designed for examining emergency and accident patienShort circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceCoolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break . the system cannot be operated normally, may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
-
Class IIofficial
Siemens Medical Solutions USA, Inc. recalls ACUSON Juniper Diagnostic Ultrasound SystemOn ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.
-
Class IIofficial
Siemens Medical Solutions USA, Inc. recalls ACUSON Maple Diagnostic Ultrasound SystemOn ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.
-
Class IIofficial
Siemens Medical Solutions USA, Inc. recalls ACUSON Redwood 1.0 Diagnostic Ultrasound SystemOn ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls Sensis Vibe HemoThe possibility of the Sensis documentation functionality application to crash.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls ARTIS icono (ceiling configuration)Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and switched on. During this time, system movement is only possible with the "override" function with very slow speed and without further collision protection. This may lead to temporary unavailability of the system for diagnostic and therapeutic usage.
-
Class IIofficial
Siemens Medical Solutions USA, Inc recalls Mobilett Mira wireless (VA20) mobile x-ray system(s)Due to the incorrect mounting of the buzzer component, there is no acoustic signal during x ray activation. Otherwise, the system is fully functional, and x ray is applied correctly. By means of the additional visual indicator (exposure LED light) the operator can monitor the x ray release. Also, after a successful x ray exposure, the clinical image is displayed on the monitor.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.