XENEX Disinfection Services Inc. recalls LightStrike Blackout Curtains
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Blackout curtains are no longer recommended as a protective barrier against UV/Ozone exposure to a bystander standing in the same room, but on the other side of the curtain from where a microbial reduction device UV cycle is taking place. Bystanders in the room during the cycle may experience temporary skin reddening, feeling of sand in eyes, nausea, headache, and/or mucous membrane irritation.
- Products
-
- LightStrike Blackout Curtains
- Company
- XENEX Disinfection Services Inc.
- Where it was sold
-
US Nationwide distribution in the states of AR, AZ, CA, CO, FL, GA, HI, ID, IL, KY, LA, MI, MN, MO, NC, ND, NV, OH, OK, OR, TN, UT, WA, WV, WY.
States: Arkansas, Arizona, California, Colorado, Florida, Georgia, Hawaii, Idaho, Illinois, Kentucky, Louisiana, Michigan, Minnesota, Missouri, North Carolina, North Dakota, Nevada, Ohio, Oklahoma, Oregon, Tennessee, Utah, Washington, West Virginia, Wyoming
- Units affected
- 377
- About this recall
- LightStrike Blackout Curtains, Part numbers; 610-00008, 610-00009, 610-00019, 610-00020, and 610-00024, used with the LightStrike Germ-Zapping Robots Models: PXUV4D, PXUV5D, LSMUV6-FT, or LSMUV6-SL, for whole room microbial reduction.