Ebi, LLC recalls Biomet SpinalPak Non-invasive Spine Fusion Stimulator System- Indicated for the treatment of an established no
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment
- Products
-
- Biomet SpinalPak Non-invasive Spine Fusion Stimulator System- Indicated for the treatment of an established no
- Company
- Ebi, LLC
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Barcodes
00812301020218- Units affected
- N/A
- About this recall
- Biomet SpinalPak Non-invasive Spine Fusion Stimulator System- Indicated for the treatment of an established nonunion acquired secondary to trauma, excluding vertebrae and all flat bones, Part Number: 1067716