Ebi, LLC recalls 48" Lead Wires
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment
- Products
-
- 48" Lead Wires
- Company
- Ebi, LLC
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Barcodes
00812301020324- Units affected
- 594 units (2 lead wires assembly)
- About this recall
- 48" Lead Wires, Replacement Part Number: 1067724-4