Olympus Corporation of the Americas recalls Brand Name: Olympus HALO PKS Cutting Forceps
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
- Products
-
- Brand Name: Olympus HALO PKS Cutting Forceps
- Where it was sold
-
Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
Nationwide
- Barcodes
00821925036390- Units affected
- 2648 units
- About this recall
- Brand Name: Olympus HALO PKS Cutting Forceps, 5mm 33cm Product Name: HALO PKS Cutting Forceps Model/Catalog Number: HACF0533 Product Description: The HALO PKS Cutting Forceps are intended to be passed through a 5 mm laparoscopic cannula. Coagulation is achieved using electrosurgical energy under laparoscopic visualization. The device is intended to be used with the Gyrus ACMI G400 Workstation only.