Bd Kiestra Lab Automation recalls BD Kiestra" ReadA; Catalog No.: 446948.
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- In certain situations, following a system reboot, one or more modules may have become unreachable via the remote connectivity interface. When this occurred, there was potential for delays in plate retrieval from the incubator while connectivity was restored. This issue was intermittent and did not affect test results, performance of the assay, or sample integrity.
- Products
-
- BD Kiestra" ReadA; Catalog No.: 446948.
- Company
- Bd Kiestra Lab Automation
- Where it was sold
-
Domestic: KS, MD, MN, NJ, TX, VA; International: Australia, Austria, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Kuwait, Mexico, Morocco, Netherlands, Norway, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, United Kingdom;
States: Kansas, Maryland, Minnesota, New Jersey, Texas, Virginia
- Barcodes
00382904469481- Units affected
- 82 units (7 US, 65 OUS)
- About this recall
- BD Kiestra" ReadA; Catalog No.: 446948.