Olympus Corporation of the Americas recalls POWERSEAL 5 mm
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Upon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to the device resulting in delay in surgery.
- Products
-
- POWERSEAL 5 mm
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of AU, CA, HK, KR.
Nationwide
- Barcodes
00821925044531- Units affected
- 160 units: (20 units US); (140 units OUS)
- About this recall
- POWERSEAL 5 mm, 23 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic. Model/Catalog Number: PS-0523CJDA