Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Zimmer, Inc. recalls ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
Products
  • ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate
Where it was sold

Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.

Nationwide

Barcodes
00889024088504
About this recall
ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Right, 6 Holes, 106 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 47-2357-017-06