Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Richard Wolf GmbH recalls PiezoWave 2 Control Unit

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The high voltage capacitor may fail early in the product life cycle after delivery of surge voltages. When this failure occurs, the capacitor may produce smoke and a burning smell. In extremely rare instances a capacitor may cause a spark or excessive heat when an internal capacitor failure occurs.
Products
  • PiezoWave 2 Control Unit
Company
Richard Wolf GmbH
Where it was sold

US State: GA

States: Georgia

Barcodes
00814334022863
Units affected
31 units
About this recall
PiezoWave 2 Control Unit, Product Number 100506US, compact myofascial acoustic compression therapy device that enables treatment with piezoelectric shockwaves (Therapeutic massager).