Richard Wolf GmbH recalls PiezoWave 2 Control Unit
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The high voltage capacitor may fail early in the product life cycle after delivery of surge voltages. When this failure occurs, the capacitor may produce smoke and a burning smell. In extremely rare instances a capacitor may cause a spark or excessive heat when an internal capacitor failure occurs.
- Products
-
- PiezoWave 2 Control Unit
- Company
- Richard Wolf GmbH
- Where it was sold
-
US State: GA
States: Georgia
- Barcodes
00814334022863- Units affected
- 31 units
- About this recall
- PiezoWave 2 Control Unit, Product Number 100506US, compact myofascial acoustic compression therapy device that enables treatment with piezoelectric shockwaves (Therapeutic massager).