ICU Medical, Inc. recalls LifeShield Drug Library Management (DLM)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Firm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate change values with certain limits, potentially resulting in over-delivery to patient. 2) An unauthorized user may modify and approve a drug library, potentially leading to incorrect program parameters being used for therapy.
- Products
-
- LifeShield Drug Library Management (DLM)
- Company
- ICU Medical, Inc.
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of CA, FL, IA, KY, LA, MA, MS, NY, PA, TN, and VA. The country of Philippines.
States: California, Florida, Iowa, Kentucky, Louisiana, Massachusetts, Mississippi, New York, Pennsylvania, Tennessee, Virginia
- Barcodes
00887709122802- Units affected
- 12 US, 3 OUS
- About this recall
- LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software SUite v2.1 and v2.2