Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medtronic Navigation, Inc. recalls Stealth S8 Clinical Software Application

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Due to a software issue, there is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation.
Products
  • Stealth S8 Clinical Software Application
Where it was sold

Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Republic Of Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, Nicaragua, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam.

Nationwide

Units affected
3286 applications.
About this recall
Stealth S8 Clinical Software Application, REF: 9735762, CE0344, Rx Only. Used in the StealthStation System as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures.