Nrt X-ray A/s recalls Intelli-C: Right side suspended Tabletop
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.
- Products
-
- Intelli-C: Right side suspended Tabletop
- Company
- Nrt X-ray A/s
- Where it was sold
-
US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, WV, LA, NY, PA, HI, NV, WI, MN, MI.
States: Hawaii, Iowa, Kentucky, Louisiana, Massachusetts, Michigan, Minnesota, North Carolina, Nevada, New York, Ohio, Pennsylvania, South Carolina, Washington, Wisconsin, West Virginia
- Units affected
- 35
- About this recall
- Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system