Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

MEDLINE INDUSTRIES, LP - Northfield recalls Medline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ADD A PACK

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.
Products
  • Medline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ADD A PACK
Where it was sold

US distribution to CA, FL, IL, IN, KY, MA, MD, MN, NY, TX, VT, WA and WI. No OUS distribution.

States: California, Florida, Illinois, Indiana, Kentucky, Massachusetts, Maryland, Minnesota, New York, Texas, Vermont, Washington, Wisconsin

Barcodes
101953275391531019532746662610195327561116101953276373921019532757557110195327635466
Units affected
1853 total
About this recall
Medline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ADD A PACK, SKU DYNJ30926O 2. CRANIOTOMY, SKU CDS983611N 3. CRANIOTOMY CDS, SKU CDS983467J 4. CRANIOTOMY PACK, SKU DYNJ85807A, DYNJ85927 5. MAJOR CRANIOTOMY PACK, SKU DYNJ82007A