Therakos Development Limited recalls Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Catalog Number: CLXUSA So
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation may cause the centrifuge bowl to dislodge, resulting in a broken bowl during the prime cycle or treatment phase, leading to a delay in patient treatment and blood loss.
- Products
-
- Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Catalog Number: CLXUSA So
- Company
- Therakos Development Limited
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Barcodes
20705030200003- Units affected
- 11,499 units
- About this recall
- Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Catalog Number: CLXUSA Software Version: Not Applicable Product Description: THERAKOS Photopheresis or extracorporeal photopheresis (ECP) is a photoimmune therapy where leukocytes are separated from whole blood via apheresis, combined with a photoactive drug (8-methoxypsoralen) and then exposed to ultraviolet A (UVA) light. All blood components, including the treated leukocytes, are returned to the patient. THERAKOS¿ Photopheresis utilizes the THERAKOS CELLEX System to combine cell separation and photoactivation into a single, closed and sterile circuit. The THERAKOS CELLEX Photopheresis System collects the buffy coat (leukocyte-enriched blood) from the patient in a discontinuous flow process and intermittently returns the remaining plasma and erythrocytes to the patient. The buffy coat is passed through the photoactivation module where the drug is activated with a precise amount of UVA light determined by the characteristics of the individual patient s buffy coat. After photoactivation, the buffy coat is immediately returned to the patient s bloodstream. Component: Not Applicable