Macherey Nagel Gmbh & Co. Kg recalls Reagent Test Strips for Blood

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The reason for the recall is one of the test pads, for the parameter leukocytes, was missing on some of the test strips. The issue was identified by a consumer by way of a Henry Schein Inc. (HSI) customer complaint. This problem can potentially result in the possibility that the parameter leucocytes (a non-specific marker for urinary track infections and an indicator for inflammatory renal processes) is interpreted instrumentally as a false negative. The following HSI item number and lot number were affected: Urispec¿ 11-Way, #102-2285, LOT 65402.
Products
  • Reagent Test Strips for Blood
Company
Macherey Nagel Gmbh & Co. Kg
Where it was sold

US Nationwide distribution in the states of CO, FL, IN, NV, TX.

States: Colorado, Florida, Indiana, Nevada, Texas

Barcodes
00304040001479
Units affected
5,200 packages
About this recall
Reagent Test Strips for Blood, Urobilinogen, Bilirubin, Protein, Nitrite, Ketones, Ascorbic Acid, Glucose, pH, Specific Gravity and Leucocytes in Urine by the Dip and Read Technique.