Macherey Nagel Gmbh & Co. Kg recalls Reagent Test Strips for Blood
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The reason for the recall is one of the test pads, for the parameter leukocytes, was missing on some of the test strips. The issue was identified by a consumer by way of a Henry Schein Inc. (HSI) customer complaint. This problem can potentially result in the possibility that the parameter leucocytes (a non-specific marker for urinary track infections and an indicator for inflammatory renal processes) is interpreted instrumentally as a false negative. The following HSI item number and lot number were affected: Urispec¿ 11-Way, #102-2285, LOT 65402.
- Products
-
- Reagent Test Strips for Blood
- Company
- Macherey Nagel Gmbh & Co. Kg
- Where it was sold
-
US Nationwide distribution in the states of CO, FL, IN, NV, TX.
States: Colorado, Florida, Indiana, Nevada, Texas
- Barcodes
00304040001479- Units affected
- 5,200 packages
- About this recall
- Reagent Test Strips for Blood, Urobilinogen, Bilirubin, Protein, Nitrite, Ketones, Ascorbic Acid, Glucose, pH, Specific Gravity and Leucocytes in Urine by the Dip and Read Technique.