Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Olympus Corporation of the Americas recalls Olympus Thunderbeat II Shears w/Ultrasonic Mode

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Potential for detachment of a distal tip component of the device during use.
Products
  • Olympus Thunderbeat II Shears w/Ultrasonic Mode
Where it was sold

International distribution in the countries of Australia, Japan, Hong Kong, and Europe.

Barcodes
04953170440007
Units affected
365 units
About this recall
Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 25cm. Model Number: TB2-0525FC.