Philips Medical Systems Nederland B.v. recalls Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722012
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.
- Products
-
- Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722012
- Where it was sold
-
Domestic: OK International: India, Spain, U.A.E.
States: Oklahoma
- Barcodes
00884838059054- Units affected
- 1 unit (US)
- About this recall
- Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722012, Software Version Number 8.1.100.