Philips Medical Systems Nederland B.v. recalls Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722012

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.
Products
  • Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722012
Where it was sold

Domestic: OK International: India, Spain, U.A.E.

States: Oklahoma

Barcodes
00884838059054
Units affected
1 unit (US)
About this recall
Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722012, Software Version Number 8.1.100.