Virtual Incision Corporation recalls MIRA SURGICAL SYSTEM
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the hand controllers and the movement of the Surgical Minibot arms.
- Products
-
- MIRA SURGICAL SYSTEM
- Company
- Virtual Incision Corporation
- Where it was sold
-
US Distribution to States: CA, FL, NE
States: California, Florida, Nebraska
- Barcodes
00850038042028- Units affected
- 3 units
- About this recall
- MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3