Healthmark Industries Co., Inc. recalls Pre-filled Water Syringe
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and concluded that the Pre-filled Water Syringe had been improperly used outside of the intended purpose.
- Products
-
- Pre-filled Water Syringe
- Where it was sold
-
US Nationwide and Internationally to countries of: Canada, Malaysia.
Nationwide
- Units affected
- 250000 units
- About this recall
- Pre-filled Water Syringe, Model Numbers SW-010-100 and SW-010