Healthmark Industries Co., Inc. recalls ChannelCheck Convenience Pack
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and concluded that the Pre-filled Water Syringe had been improperly used outside of the intended purpose.
- Products
-
- ChannelCheck Convenience Pack
- Where it was sold
-
US Nationwide and Internationally to countries of: Canada, Malaysia.
Nationwide
- Units affected
- 193035 units
- About this recall
- ChannelCheck Convenience Pack, Model Numbers UCC-222CP and UCC-222-ESP (Sample)