Covidien recalls Newport HT70 Ventilator
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown alert alarm fails to alarm effectively during shut down, which may result in respiratory failure, hypoventilation, low oxygen saturation, hypoxia, treatment delay.
- Products
-
- Newport HT70 Ventilator
- Company
- Covidien
- Where it was sold
-
Worldwide - US Nationwide distribution including in the states of MA, CT, KS, OH, PR, NH, GA, WI, MS, VA, CO, TX, MN, AL, IA, NY, UT, NM, KY, RI, SD, PA, IA, CA, MO, HI, MI, IL, MT, NJ, TN, SC, FL, IN, AZ, ID, LA, WA, NC, MD, WY, NE, NV, ND, ME, AK, DE, OK, OR and the countries of Japan, Canada, Singapore, Lithuania, Kazakhstan, Slovenia, Colombia, China, Qatar, United Arab Emirates, Greece, Saudi Arabia, Nigeria, Romania, Netherlands, Lebanon, Oman, Jordan, Malawi, India, Croatia, Portugal, Turkey, Brazil, Pakistan, Trinidad and Tobago, Slovakia, Chile, Egypt, Switzerland, South Africa, Thailand, Bangladesh, Myanmar, Peru, Uruguay, Hungary, Argentina, Iran, Islamic Republic Of, Costa Rica, Vietnam, Australia, Guatemala, Mexico, Guyana, Jamaica, Dominican Republic, Bahamas, Bosnia and Herzegovina, South Korea, Tunisia, Ukraine, Uzbekistan, Indonesia, Burkina Faso, Italy, Panama, Sri Lanka, Malaysia, Namibia, Swaziland, Spain, Philippines, Bhutan, Mauritius, Poland, France, Cambodia, Czechia, Israel, Norway, Russian Federation, United Kingdom, Canary Islands, Kenya, Czech Republic, Croatia , Estonia, Maldova, Palestine.
Nationwide
- Units affected
- 5394
- About this recall
- Newport HT70 Ventilator, REF: HT70M-JP-NA, HT70M-CN-NA, HT70M-ES-EU, HT70M-PT-BR, HT70M-SY-EU, HT70M-WW-EU, HT70M-WW-NA, HT70-SY-EU. Newport HT70Plus Ventilator, REF: HT70PM-ES-EU, HT70PM-ES-NA, HT70PM-JP-NA, HT70PM-PT-BR, HT70PM-SY-AS, DLHT70PM-WW-NA, HT70PM-SY-NA, HT70PM-SY-UK, HT70PM-WW-EU, HT70PM-WW-UK, HT70PM-SY-EU, HT70PM-WW-NA. HT70-2 CONTROL BOARD ROHS, REF: GR-PCB3252A. HT70-2 CONTROL BOARD ROHS X1, REF: GR105796. HT70P CONTROL BOARD ROHS X1, REF: GR105795. HT70 Conversion Kit, REF: GRFSKHT70M-2.