Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Newport Corp recalls Newport s N-LHP-928(j) units if sold as a laser head/power supply set

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
According to 21 CFR 1040.10(f)(4), each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. However, the key of the subject laser products can be removed from key switch and the laser remains on. This is a safety concern.
Products
  • Newport s N-LHP-928(j) units if sold as a laser head/power supply set
Company
Newport Corp
Where it was sold

US Nationwide

Nationwide

Units affected
19
About this recall