Newport Corp recalls Newport s N-LHP-928(j) units if sold as a laser head/power supply set
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- According to 21 CFR 1040.10(f)(4), each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. However, the key of the subject laser products can be removed from key switch and the laser remains on. This is a safety concern.
- Products
-
- Newport s N-LHP-928(j) units if sold as a laser head/power supply set
- Company
- Newport Corp
- Where it was sold
-
US Nationwide
Nationwide
- Units affected
- 19
- About this recall
- Newport s N-LHP-928(j) units if sold as a laser head/power supply set, and N5-LPL-944-080 and N6-LPL-944-080 if sold as a power supply alone.