Turncare, Inc recalls Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Softwa
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- A software issue where "low" and "terminal" battery alerts did not display on the screen to the user in specific circumstances when the device was not running patient therapy.
- Products
-
- Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Softwa
- Company
- Turncare, Inc
- Where it was sold
-
US Nationwide distribution in the states of Texas, Illinois, Pennsylvania, Florida, Connecticut
States: Connecticut, Florida, Illinois, Pennsylvania, Texas
- Barcodes
00860001236453- Units affected
- 1081 units
- About this recall
- Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Software Version: 2.5.4-59 (Fix)