Turncare, Inc recalls Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Softwa
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- A software update corrected an issue where "low" and "terminal" battery alerts did not display on the screen to the user in specific circumstances when the device was not running patient therapy.
- Products
-
- Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Softwa
- Company
- Turncare, Inc
- Where it was sold
-
US Nationwide distribution in the states of Texas, Illinois, Pennsylvania, Florida, Connecticut.
States: Connecticut, Florida, Illinois, Pennsylvania, Texas
- Barcodes
00860001236453- Units affected
- 1142 units
- About this recall
- Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Software Version: 2.5.4-69 (Fix)