Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Orthofix Srl recalls REF: 99-709245

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Retrospective; Their is the potential that implantable intramedullary lengthening nails may bend at the distal tail during insertion resulting in having to be replaced.
Products
  • REF: 99-709245
Company
Orthofix Srl
Where it was sold

US: Unknown, OUS: Unknown

Barcodes
18059015373653
Units affected
2 units
About this recall
REF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, Sterile