Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

ICU Medical, Inc. recalls ICU Medical Plum Solos

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight or BSA (body surface area) based medication rulesets. The limit can be configured for Dose, Loading Dose, and Bolus Dose.
Products
  • ICU Medical Plum Solos
Where it was sold

Worldwide - US Nationwide distribution in the states of FL, KY, VA, MS, MA, PA, NY, CA, IA and the country of Philippines.

States: California, Florida, Iowa, Kentucky, Massachusetts, Mississippi, New York, Pennsylvania, Virginia

Units affected
214 units
About this recall
ICU Medical Plum Solos, Item number 400011001; infusion pump