Olympus Corporation of the Americas recalls OES 4000 Hysteroscope

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.
Products
  • OES 4000 Hysteroscope
Where it was sold

US Nationwide distribution in the state of NY.

States: New York

Barcodes
04042761006361
Units affected
1 unit
About this recall
OES 4000 Hysteroscope, Model/Catalog Number: A4674A