Olympus Corporation of the Americas recalls OES 4000 Hysteroscope
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.
- Products
-
- OES 4000 Hysteroscope
- Where it was sold
-
US Nationwide distribution in the state of NY.
States: New York
- Barcodes
04042761006361- Units affected
- 1 unit
- About this recall
- OES 4000 Hysteroscope, Model/Catalog Number: A4674A