Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Covidien, LLC recalls EsoFLIP

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Due to saline conductivity issues, their is a potential that dilation catheters may provide inaccurate esophageal diameter measurements that may lead to various harms.
Products
  • EsoFLIP
Where it was sold

Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DC, FL, GA, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Canada, Denmark, Germany, Ireland, Italy, Kuwait, New Zealand, Norway, Portugal, Qatar, South Africa, Switzerland, United Arab Emirates.

Nationwide

Barcodes
1088452180945120884521809458
Units affected
N/A
About this recall
EsoFLIP, 30mm BALLOON DILATION CATHETER, REF ES-330, Rx Only