Boston Scientific Corporation recalls LUX-Dx II Plus
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
- Products
-
- LUX-Dx II Plus
- Where it was sold
-
Worldwide distribution - US Nationwide and the country of Belgium.
Nationwide
- Barcodes
008025266207060080252661382100802526613814- Units affected
- 148 units
- About this recall
- LUX-Dx II Plus, Arrythmia detector and alarm, Model M312, with SERVER SW LATITUDE DRAGON EU (Europe), Model 6446, Version 7.5, or SERVER SW LATITUDE DRAGON AU (Australia/New Zealand), Model 6448, Version 7.5