Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Boston Scientific Corporation recalls VersaCross Transseptal Sheath Kit (VXS) Product Description/IDs VXS0201 VXS 63CM/45DEG/D0 SHEATH VXS2112 VERSA

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
Products
  • VersaCross Transseptal Sheath Kit (VXS) Product Description/IDs VXS0201 VXS 63CM/45DEG/D0 SHEATH VXS2112 VERSA
Where it was sold

Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.

Nationwide

Barcodes
0068544702030100685447019251
Units affected
14,724
About this recall
VersaCross Transseptal Sheath Kit (VXS) Product Description/IDs VXS0201 VXS 63CM/45DEG/D0 SHEATH VXS2112 VERSACROSS SHEATH (63CM 45DEG D0)-J Used for the percutaneous introduction of various types of cardiovascular catheters and guidewires to all heart chambers, including the left atrium via transseptal perforation / puncture.