ICU Medical, Inc. recalls SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which may result in inadequate disinfection.
- Products
-
- SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve
- Company
- ICU Medical, Inc.
- Where it was sold
-
Worldwide - US Nationwide distribution including in the states of TX, SD, ID, LA, IN, CO, OK, MO, MN, OR, AZ, NC, TN, VA, CA, IL, FL, MA, OH, MD, PA, NJ, NY, WA, GA, UT, MI, MS, AR, AL, NV, KS, IA, WV, HI, KY, NM and the country of CA.
Nationwide
- Units affected
- 1,445,600
- About this recall
- SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF: SCXT3-2000