Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

ICU Medical, Inc. recalls SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which may result in inadequate disinfection.
Products
  • SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve
Where it was sold

Worldwide - US Nationwide distribution including in the states of TX, SD, ID, LA, IN, CO, OK, MO, MN, OR, AZ, NC, TN, VA, CA, IL, FL, MA, OH, MD, PA, NJ, NY, WA, GA, UT, MI, MS, AR, AL, NV, KS, IA, WV, HI, KY, NM and the country of CA.

Nationwide

Units affected
1,445,600
About this recall
SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF: SCXT3-2000