Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

CooperSurgical, Inc. recalls Endosee System Convenience Kit. Model Number: ES-TRAY. Endosee System Convenience Kit

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the peel away lid labels incorrectly identify the iodine pouches as sterile.
Products
  • Endosee System Convenience Kit. Model Number: ES-TRAY. Endosee System Convenience Kit
Where it was sold

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Washington DC.

Nationwide

Barcodes
60888937027268
Units affected
991 units
About this recall
Endosee System Convenience Kit. Model Number: ES-TRAY. Endosee System Convenience Kit