CooperSurgical, Inc. recalls INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.
- Products
-
- INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP
- Company
- CooperSurgical, Inc.
- Where it was sold
-
Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.
Nationwide
- Barcodes
2088893700943360888937009448208889370094572088893700946420888937009471- Units affected
- 2,643 boxes (13,215 units)
- About this recall
- INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-2707, 44-2709, 44-2710, 44-2712, 44-2715.