Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Ethicon, Inc. recalls VICRYL Plus Antibacterial (polyglactin 910) Suture - Indicated for use in general soft tissue approximation an

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
Products
  • VICRYL Plus Antibacterial (polyglactin 910) Suture - Indicated for use in general soft tissue approximation an
Where it was sold

Worldwide distribution - US Nationwide including PR and the countries of AR, AU, AU, BE, BR, CA, CA, CL, CN, CO, CR, EC, IL, JP, JP, KW, MX, NZ, PA, PR, PR, SA, SG, UY, ZA.

Nationwide

Barcodes
10705031051546
Units affected
55296 units
About this recall
VICRYL Plus Antibacterial (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological tissues. PRODUCT CODE: VCP196H VCP417H VCP602H VCP603H VCP662H VCPB259H VCPB946H