Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Inspire Medical Systems Inc. recalls Inspire Model 3028

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
There is a potential manufacturing defect which can lead to electrical leakage in the sensing circuit which causes a system malfunction and requires a revision surgery to replace the IPG,
Products
  • Inspire Model 3028
Where it was sold

Worldwide distribution - US Nationwide and the country of Germany.

Nationwide

Barcodes
10855728005847
Units affected
24 devices
About this recall
Inspire Model 3028, IV Implantable Pulse Generator