Inspire Medical Systems Inc. recalls Brand Name: Inspire Medical Systems
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.
- Products
-
- Brand Name: Inspire Medical Systems
- Company
- Inspire Medical Systems Inc.
- Where it was sold
-
Worldwide distribution - US Nationwide and the country of Hong Kong.
Nationwide
- Units affected
- 50 impacted units
- About this recall
- Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The Inspire¿ Respiratory Sensing Lead (Model 4340) is designed to detect respiratory effort. The lead features a pressure sensitive membrane that converts the mechanical energy of respiration into an electrical signal. The lead incorporates a standard connector for coupling to the implantable pulse generator (IPG). Component: N/A