Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Horiba Instruments Incorporated recalls HORIBA custom configured fluorescence instrument

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
HORIBA custom configured fluorescence instrument has a laser safety defect. This custom configuration of the modular Instrument removed the primary light source and the excitation monochromator. A plastic cap was installed to plug the optical port, where a more durable beam stop should have been used. The defect may allow the class 4 laser beam to burn (or transmit) through the plastic cap to exit the protective enclosure.
Products
  • HORIBA custom configured fluorescence instrument
Company
Horiba Instruments Incorporated
Where it was sold

US Nationwide and Worldwide Distribution

Nationwide

Units affected
4 units
About this recall
HORIBA custom configured fluorescence instrument, modular Fluorolog-QM