Philips Medical Systems Nederland B.v. recalls Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.
- Products
-
- Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221
- Where it was sold
-
Worldwide distribution - US Nationwide and the country of Canada.
Nationwide
- Units affected
- 1,718 units
- About this recall
- Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230), Azurion 5 (722281), Azurion 5 M12 (722227, 722231), Azurion 5 M20 (722228, 722232, 722281), Azurion 7 (722282), Azurion 7 B12 (722225, 722235), Azurion 7 B20 (722068, 722226, 722236), Azurion 7 M12 (722078, 722223, 722233), Azurion 7 M20 (722079, 722224, 722234, 722282).