Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Sight Diagnostics Ltd recalls Sight OLO CBC Test Kit

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
One Test Kit lot (TK1-10114) shows a positive bias in the platelet count due to bacterial contamination which can result in elevated counts with a bias of 7.5 k/ul. This results in the test kit lot performing outside of the device specification.
Products
  • Sight OLO CBC Test Kit
Company
Sight Diagnostics Ltd
Where it was sold

Worldwide - US Nationwide distribution in the states of AL, FL, MS, TX, WI and the countries of Canada, Israel, Slovenia, Trinidad & Tobago.

States: Alabama, Florida, Mississippi, Texas, Wisconsin

Units affected
7,450 units (2,800 US, 4,650 OUS)
About this recall
Sight OLO CBC Test Kit, Model/Catalog Number: TK1. A disposable cartridge and sample preparation tools for use with the Sight OLO CBC analyzer.