Sight Diagnostics Ltd recalls Sight OLO CBC Test Kit
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- One Test Kit lot (TK1-10114) shows a positive bias in the platelet count due to bacterial contamination which can result in elevated counts with a bias of 7.5 k/ul. This results in the test kit lot performing outside of the device specification.
- Products
-
- Sight OLO CBC Test Kit
- Company
- Sight Diagnostics Ltd
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of AL, FL, MS, TX, WI and the countries of Canada, Israel, Slovenia, Trinidad & Tobago.
States: Alabama, Florida, Mississippi, Texas, Wisconsin
- Units affected
- 7,450 units (2,800 US, 4,650 OUS)
- About this recall
- Sight OLO CBC Test Kit, Model/Catalog Number: TK1. A disposable cartridge and sample preparation tools for use with the Sight OLO CBC analyzer.