Maquet Cardiovascular, LLC recalls The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting System

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.
Products
  • The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting System
Where it was sold

Worldwide distribution - US Nationwide and the countries of Australia Brazil, Canada, China, EEA, India, Korea, New Zealand, Southern Africa, Taiwan, Thailand, Turkey, & United Kingdom.

Nationwide

Barcodes
0060756770040600607567700451
Units affected
430,037 units
About this recall
The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting System, labeled as the following: 1. VasoView Hemopro 2, Model Number: VH-4000. 2. VasoView Hemopro 2 with Vasoshield. Model Number: VH-4001. The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting Systems are designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.