MEDLINE INDUSTRIES, LP - Northfield recalls SwabFlush Prefilled Syringe with ICU Medical SwabCaps
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- SwabFlush syringes are affected by the ICU Medical recall of their SwabCap Disinfecting Cap due to a manufacturing issue which affects the seal between the foil lid and plastic container of the Swab Cap. The issue may result in a reduced level of disinfection of the luer access valve, potentially resulting in patient infection/sepsis.
- Products
-
- SwabFlush Prefilled Syringe with ICU Medical SwabCaps
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Canada and Panama.
Nationwide
- Barcodes
00363807103016- Units affected
- 1,427,100 eaches
- About this recall
- SwabFlush Prefilled Syringe with ICU Medical SwabCaps, 10mL Model no. EMZE010301