Mermaid Medical A/S recalls The D*Clot¿ HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. T
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Internal product testing failed bubble leak testing (ASTM F2096), which indicates the sterility barrier may not be fully intact.
- Products
-
- The D*Clot¿ HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. T
- Company
- Mermaid Medical A/S
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of OH, MO, CO, NC, TX, MA, GA, FL, IL, IN, MD, PA, VA, and the country of Chile.
States: Colorado, Florida, Georgia, Illinois, Indiana, Massachusetts, Maryland, Missouri, North Carolina, Ohio, Pennsylvania, Texas, Virginia
- Units affected
- 190 devices
- About this recall
- The D*Clot¿ HD Rotational Thrombectomy System is a percutaneous catheter-based system available in 6 French. The disposable system consists of an outer sheath through which a rotatable radiopaque spiral shaft extends ending in an eccentric blunt distal tip. Once activated, the shaft rotates at approximately 10.000 RPMs to break thrombus into smaller particles, which can be simultaneously aspirated using a dedicated vacuum syringe. The catheter is a stand-alone system and is powered by a 4.5 VDC alkaline battery pack in the control unit.