Thoratec Switzerland GMBH recalls 2nd Generation CentriMag Primary Console
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Due to a component change, the console part of a circulatory support system may not conform to the IEC 61000-4-5 standard, so if subjected to a power surge above 1.8kV and up to 2.0 kV, the console may shut down completely without alarm, pumping would cease, and the patient would have to be switched to a backup circulatory support system.
- Products
-
- 2nd Generation CentriMag Primary Console
- Company
- Thoratec Switzerland GMBH
- Where it was sold
-
Worldwide - US Nationwide distribution included in the states of CA, PA, FL, IL, NY, AR, GA, WI, AZ, TN, TX, NC, MT, VA, MA, NJ, NE, WV, ME, OK, CO, OH, KC, WA, IN, CT, MI, OR, ID, MO, SC, KY, LA, NM, NV, VT, IA, MD and the countries of ALGERIA, AUSTRALIA, AUSTRIA, BANGLADESH, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRAN, IRAQ, ISRAEL, ITALY, KENYA, KOREA, KUWAIT, LITHUANIA, MACEDONIA MALAYSIA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NORWAY, OMAN, PERU, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, UNITED ARAB EMIRATES, UNITED KINGDOM.
States: Arkansas, Arizona, California, Colorado, Connecticut, Florida, Georgia, Iowa, Idaho, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Missouri, Montana, North Carolina, Nebraska, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, Tennessee, Texas, Virginia, Vermont, Washington, Wisconsin, West Virginia
- Barcodes
07640135140689- Units affected
- 3,425
- About this recall
- 2nd Generation CentriMag Primary Console, REF: 201-90401, 201-90701, 201-90411, 201-90421, 201-30300, L201-90401, L201-90411, L201-90421, part of the CentriMag Circulatory Support System