Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Siemens AG/Siemens Healthcare GmbH recalls ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and int

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.
Products
  • ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and int
Where it was sold

Worldwide - US Nationwide distribution in the states of TX, MA, NH, MN, OK, NJ, SC, NY, MI, IL, MO, NC.

States: Illinois, Massachusetts, Michigan, Minnesota, Missouri, North Carolina, New Hampshire, New Jersey, New York, Oklahoma, South Carolina, Texas

Units affected
201 units in total
About this recall
ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.