Siemens AG/Siemens Healthcare GmbH recalls ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and int
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.
- Products
-
- ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and int
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of TX, MA, NH, MN, OK, NJ, SC, NY, MI, IL, MO, NC.
States: Illinois, Massachusetts, Michigan, Minnesota, Missouri, North Carolina, New Hampshire, New Jersey, New York, Oklahoma, South Carolina, Texas
- Units affected
- 201 units in total
- About this recall
- ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.