Nalu Medical, Inc. recalls Neurostimulation Kit (Ported
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Neurostimulation Systems have a manufacturing non-conformance that may reduce the efficiency of the radiofrequency signals between the Implantable Pulse Generator (IPG) and the Therapy Disc, causing the IPG to be non-functional, which causes the device to be non-functional so it will no longer inhibit the transmission of pain signals to the brain.
- Products
-
- Neurostimulation Kit (Ported
- Company
- Nalu Medical, Inc.
- Where it was sold
-
US Nationwide distribution in the states of MO, IL, CA, TX, TN, OK, AZ, AR.
States: Arkansas, Arizona, California, Illinois, Missouri, Oklahoma, Tennessee, Texas
- Units affected
- 13
- About this recall
- Neurostimulation Kit (Ported, Dual 8, 40 cm), REF: 71005-US Peripheral Nerve Stimulation Kit (Ported, Dual 8, 40 cm), REF: 71019-US Components of the Nalu Spinal Cord Stimulation System.