Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Philips Medical Systems Nederland B.v. recalls Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
Products
  • Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063
Where it was sold

Worldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussal, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Greece, Guinea, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen.

Nationwide

Barcodes
0088483808527500884838099203008848380852820088483809921000884838099227008848380992340088483808535000884838099265008848380853670088483809927200884838085251008848380992410088483808526800884838099258
Units affected
3069 units (872 US, 2224 OUS)
About this recall
Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063, 722221), Azurion 3 M15 (722064, 722222), Azurion 5 M12 (722227), Azurion 5 M20 (722228), Azurion 7 B12 (722067, 722225), Azurion 7 B20 (722068, 722226), Azurion 7 M12 (722078, 722223), Azurion 7 M20 (722079, 722224).